EN 50147-1 applies to all types of shielded enclosures, from small equipment rooms to large anechoic chambers. Its core technical specifications include:

Before formal validation, leakage tests are performed around critical "weak points" like door frames, air ducts, filters, and joints. Acceptance measurements are typically conducted before the installation of internal RF absorbers to ensure the base metallic structure is sound.

Loop antennas are placed parallel to the wall, directly opposite each other. en 50147-1 pdf

The standard is the European benchmark for measuring the shielding attenuation of anechoic chambers and shielded enclosures. Published by CENELEC, it provides a rigorous, standardized procedure to verify that these rooms provide the electromagnetic isolation necessary for high-accuracy EMC testing . Scope and Applicability

Measurements are mandated at frequencies including 100 MHz, 400 MHz, 1 GHz, 10 GHz, and 40 GHz. Essential Test Equipment EN 50147-1 applies to all types of shielded

To comply with EN 50147-1, laboratories must use instrumentation with high dynamic range and stability:

Specialized sets (loops, dipoles, biconical, and horn) calibrated for their respective frequency bands. Importance for EMC Compliance Loop antennas are placed parallel to the wall,

Covers a broad spectrum from 9 kHz to 40 GHz .